Honeywell Analytics XD MicroDock II (HA-DOCK2) Base Only |
Model #: HA-DOCK2
| |
|
|
|
Non-stocked item. Please allow 7 to 10 days for delivery.
The Honeywell MicroDock II test/calibration docking system packs advanced
technology within a compact design to offer the industry's foremost automatic
calibration, testing, and data management system for portable gas detection
equipment. Compatible with the portable Honeywell XD and X5 gas detectors, the
Honeywell MicroDock II test/calibration docking system expands up to 10 docking
modules (including up to 6 charging units).
Fully portable, lightweight, and self-contained, the Honeywell MicroDock II
test/calibration docking system requires no computer. Instrument records are
stored for easy retrieval in the MicroDock II and each individual gas detector.
Each Honeywell MicroDock II test/calibration docking system can be connected to
a LAN. Cost and time requirements for routine maintenance are reduced through
sophisticated automatic calibration and test procedures.
The Honeywell MicroDock II test/calibration docking system includes Honeywell's
Fleet Manager software, which consolidates information to a central database to
take the work out of data analysis and management. This software tool enables
users to:
- Create reports automatically
- Sort, format and export data
- View history
- Archive easily
- Customize with user-definable fields
User-friendly, flexible, and expandable, the Honeywell MicroDock II
test/calibration docking system provides the ultimate solution for managing
high-performance, portable Honeywell XD and X5 gas detectors.
- Manages calibration, bump testing and data management of up to ten
Honeywell Analytics’ X5 and XD portable gas detectors and recharging of up
to six X5 units
- Automatic “hands-free” calibration, bump test and alarm check
- Portable, self-contained module – can operate independently or connect
to a PC with USB interface
- Flexible operation via line power or set of four C-cell batteries
- Expandable utility - add up to 10 additional docking modules to system
via simple, plug-in connections – no external pressure transducers or gas
lines necessary
- Multi-language support (English, French, German, Spanish & Portuguese)
- Included Fleet Manager software consolidates information into a central
database; creates reports automatically
Specifications:
- Power Supply: 6V wall adapter or four
C-cell batteries
- Real-time clock: Provides time and date
stamp for “last calibration” and “last bump” data
- Data storage: Automatic (instrument and
base station)
- 32 MB MultiMediaCard data storage system
- External interface: USB 2.0 interface for
PC (USB 2.0 full speed)
- Pump: DC motor, micro-diaphragm; 3.3V PCB
mount
- Flow rate: Typically 300 mL/min
- Calibration gas cylinder inputs 2-gas
inlets (standard)
- 5-gas inlets (maximum)
- Automatic tests: Functional bump,
calibration, audible alarm, visual alarm
- Configuration recognition: Automatic
(instrument and sensor)
- Detector configuration: User-settable with
Fleet Manager applications
- Calibration gas connections: Built-in
(base station)
- 1/8” SMC connect sub-miniature coupling
- Solenoid: Built-in (docking modules)
- LED INDICATORS: Yellow – TEST (on each
docking module); Green - PASS; Red - FAIL
- LCD: 4 line x 16 characters, wide viewing
angle, user-selectable backlighting
- Dimensions: 8.3" x 10.4" x 3.2"
(21.2x26.3x8.2cm) (base station plus one docking module)
- Enclosure: Impact- resistant PC/ABS
(polycarbonate)
- COMMAND KEYS: Base station Menu
navigation, Docking modules One touch Bump-test initiation, One touch
Calibration initiation, One touch data transfer on XD
- Communications Method: Infrared
communication between docking module and detector
- Limited 2 year manufacturers warranty
Download Users Manual (PDF format)
 |
|
 |
|
|
|
|
|
|
|
|