Template for UV Lamp Intensity Verification Using UV Light Meter

Here's a concise yet comprehensive SOP (Standard Operating Procedure) template designed specifically for verifying the UV bulbs using a UV light meter, aligned with Standard Methods 9020-2015 and Standard Methods 9223-2016. This can be customized by your instrumentation staff as required.


Standard Operating Procedure (SOP)

UV Lamp Intensity Verification Using UV Light Meter

(Reference: Standard Methods 9020-2015, Standard Methods 9223-2016)


1. Purpose

To describe the method used to verify the intensity/wattage output of UV bulbs utilized in handheld bacteriological testing units, ensuring compliance with ELAP laboratory accreditation requirements.


2. Scope

This SOP applies to personnel responsible for verifying the intensity/wattage of UV bulbs used in bacteriological test methods, such as coliform detection, according to the referenced standard methods.


3. Definitions


4. Equipment and Supplies


5. Procedure

5.1 Preparation

5.2 UV Intensity Measurement

  1. Switch on the UV lamp and allow it to stabilize for at least 5 minutes before measurement.
  2. Place the UV meter sensor directly beneath the UV bulb at a fixed, documented distance (e.g., 5 cm or per manufacturer specification).
  3. Record the intensity reading from the UV meter in µW/cm² or mW/cm².
  4. Perform the measurement at least three times, recording each reading on the provided log sheet. Take the average reading for accuracy.

5.3 Acceptance Criteria


6. Quality Control


7. Documentation and Records

Maintain a UV verification log for all UV bulbs including:

Date Instrument ID UV Meter ID Distance (cm) Intensity Reading (µW/cm² or mW/cm²) Average Intensity Acceptance (Pass/Fail) Analyst Initials
MM/DD/YYYY Reading 1:
MM/DD/YYYY Reading 2:
MM/DD/YYYY Reading 3:

8. References


9. Review and Approval

Prepared by (Instrumentation Staff): Date:
Reviewed by (Lab Supervisor or QA Officer): Date:
Approved by (Lab Director or QA Manager): Date:

Adapt this template to your laboratory’s specific needs and instrumentation.