HD750 Low Range High Resolution Gauge or Differential Pressure Manometer.
- 11 selectable units of measure
- Max/Min/Avg recording with elapsed timer and relative time stamp
- Data Hold functions
- Large LCD display with back-lighting
- Zero function for offset correction or measurement
- Built-in USB port (software and cable included)
- Auto power off after 20 minutes (can be disabled)
- Low battery indication
- Threaded tripod mount
Specifications:
|
|
Range |
Resolution |
| in water |
±138.30 |
0.01 |
| psi |
±5.000 |
0.001 |
| mbar |
±344.7 |
0.1 |
| kPa |
±34.47 |
0.01 |
| in Hg |
±10.18 |
0.001 |
| mm Hg |
±258.5 |
0.1 |
| oz cubic
inches |
±80.00 |
0.01 |
| ft water |
±11.530 |
0.001 |
| cm water |
±350.1 |
0.1 |
| kg cm2 |
±0.351 |
0.001 |
| bar |
±0.344 |
0.001 |
- Accuracy: ±0.3% FS + 1 digit
- Repeatability: ±0.2% to 0.5% FS
- Linearity/Hysteresis: ±0.3%FS
- Maximum pressure: 10psi
- Sample rate: 0.5 seconds typical
- Compatibility: air or non-corrosive and non-ionized gas (not intended
for liquids)
- Input Ports: 2 metal 4.8mm, barbed for 1/8" (3.17mm) ID tubing
- Display: four digit LCD with Function Indicators
- Battery power: (1) 9v battery (included)
- Battery life: 125 hours typical
- Operating temperature range: 32° to 122° F (0° to 50° C)
- Operating humidity range: < 80% RH
- Storage temperature range: 14° to 140° F (0° to 60° C)
- Storage humidity range: < 80% RH
- Dimensions: 8.2 x 2.9 x1.9" (210 x 75 x 50mm)
- Weight: 9.8 (280g)
Supplied with:
- Instruction manual
- Windows compatible data acquisition software
- USB to mini-USB interconnect cable
- 9V alkaline battery
- (2) sets of connection hoses (12" each)
- Padded carrying case