Lutron SP-82UV Pocket UV Light Meter, Wide Range, 250 to 390 nm, Auto Ranging
The Lutron SP-82UV measures ultraviolet irradiance in a pocket sized instrument. The sensor is built into the body, the meter auto ranges across the whole span from 1 µW/cm² up to 20 mW/cm², and the "bone" shaped case with its wristlet is made to be worked one handed — held out into the light while the display is read.
The sensor covers a band pass of 250 to 390 nm with a cosine correction filter, so light arriving at an angle is weighted the way a flat surface actually receives it. Calibration is performed under UVA light against a standard UVA meter.
Readings are frozen with the hold button, and a record function captures maximum and minimum with recall. The display reverses on command so the meter can be read comfortably whichever way it is being held, and a backlight makes it usable in a darkened room or shaded space.
Ultraviolet light injures eyes and skin, and the damage gives no warning while it is happening. Never look at an energized UV lamp, and never look at the sun. Wear UV rated eye protection and cover exposed skin when working near a UV source.
NOTE: This is not a germicidal meter. Although the sensor band nominally begins at 250 nm, the instrument is calibrated under UVA light against a UVA standard, so a reading taken from a 254 nm germicidal lamp cannot be relied on. For verifying disinfection output, use a dedicated UVC meter such as the UVC-254SD.
Typical Applications
Outdoor and general:
- Measuring the sun's ultraviolet output, and how it varies with time of day, season and cloud cover.
- Checking the ultraviolet transmission of window glass, film, tint, skylights and shade structures.
- Verifying the performance of UV blocking fabrics, films and protective equipment.
- Occupational exposure surveys for outdoor workers.
Industrial:
- Curing of inks, adhesives and coatings — verifying lamp output before a run and tracking it as lamps age.
- Photoresist exposure and graphic arts platemaking.
- Photochemical matching.
- EPROM erasure equipment.
- Monitoring ultraviolet radiation hazards around welding operations.
- Blacklight and inspection lamp output checks, including magnetic particle and dye penetrant inspection.
- Tanning equipment output verification.
Laboratory:
- Weathering and degradation studies.
- Virology and microbial genetics.
- DNA research and transilluminator output checks.
- Biological hoods fitted with UVA sources.
- General laboratory use.
Features
- Ultraviolet light measurement. Calibration is executed under UVA light and compared against a standard UVA light meter.
- Wide ultraviolet sensor spectrum, band pass 250 nm to 390 nm.
- Exclusive ultraviolet photo sensor with a cosine correction filter.
- Auto ranging across the full span, from 1 µW/cm² resolution up to 20 mW/cm².
- Tiny bone shaped case, light and small enough to work one handed.
- Wristlet gives the instrument extra protection during one handed operation.
- Built-in microprocessor circuit assures excellent performance and accuracy.
- Concise, compact button arrangement for easy operation.
- Records the maximum and minimum values with recall.
- Hold function freezes the current reading.
- Reversible display direction, so the meter reads right way up whichever way it is held.
- LCD backlight for use in shaded or darkened areas.
- Auto power off after approximately 10 minutes, which can be disabled by selecting the record function.
- Runs from a single CR2032 coin cell.
Specifications
- Measurement: ultraviolet light irradiance.
- Sensor spectrum: band pass 250 nm to 390 nm.
- Sensor: exclusive ultraviolet photo sensor with cosine correction filter.
- Low range: 0 to 1999 µW/cm², 1 µW/cm² resolution.
- High range: 2 to 20.00 mW/cm², 0.01 mW/cm² resolution.
- Ranging: automatic.
- Accuracy: ± (4% F.S. + 2 digits), at 73 ± 9 °F (23 ± 5 °C).
- Display: 8 mm LCD, with selectable reversed reading direction and a backlight.
- Over input indication: display shows "- - - -" above 20 mW/cm².
- Controls: power, hold and REC buttons.
- Memory: records maximum and minimum readings with recall.
- Power off: manual by holding the power button, or automatic after approximately 10 minutes.
- Dimensions: 4.7 x 1.8 x 0.8" (120 x 45 x 20 mm), height x width x depth.
- Weight: 5.6 oz (160 g), battery included.
- Operating temperature: 32 to 122 °F (0 to 50 °C). NOTE most batteries are not rated for use below -4 °F or above 125 °F.
- Operating humidity: maximum 80% RH, non-condensing.
- Power supply: one CR2032 3V lithium coin cell.
- Power consumption: approximately 5 mA DC.
Sensor Care
The sensor window is part of the measurement. Keep it clean and free of fingerprints, dust and moisture, and wipe it only with a soft dry cloth. Store the meter dry and out of direct sunlight. A dirty or clouded window reads low, and no calibration can correct for it.
Accessories Included
- Wristlet.
- CR2032 battery.
- Instruction manual.